Oxygen analyser readings: Humidity, altitude and temperature

An oxygen analyser doesn’t measure the percentage of oxygen. Its sensor responds to the partial pressure of oxygen, so humidity, altitude, the weather and temperature all move the reading. Calibrate to 20.9 % in humid air, or at home and then use the analyser at a dive site in the mountains, and the same gas shows a different figure.

Enter the gas and the conditions at calibration and at measurement, and the calculator estimates what the analyser should show. Or enter what it shows to find out how much oxygen the gas really holds.

Background

What the cell measures

The cell in a nitrox analyser produces a current in proportion to the partial pressure of oxygen at its membrane, not to the percentage. At calibration the analyser is told which percentage that current means; afterwards it turns the current back into a percentage on that basis.

p_{O_2} = x \cdot (p - p_w)
p_{O_2}Partial pressure of oxygen in hPa
xOxygen fraction of the dry gas
pAmbient pressure (absolute) in hPa
p_wPartial pressure of water vapour in hPa

The reading

D = S \cdot \frac{x_m \cdot (1 - h_m) \cdot p_m}{x_c \cdot (1 - h_c) \cdot p_c} \cdot G
DDisplayed reading in %
SValue set at calibration in %
x_mOxygen fraction of the dry gas measured
h_mWater-vapour fraction of the gas measured
p_mAmbient pressure at measurement (absolute) in hPa
x_cOxygen fraction of the dry calibration gas (0.20946 for air)
h_cWater-vapour fraction of the calibration air
p_cAmbient pressure at calibration (absolute) in hPa
GTemperature factor of the cell (1 for a compensated cell)

An analyser with a built-in barometer corrects for the air pressure itself; then p_m and p_c drop out. For ‘What is really in it?’ the same formula is solved for x_m.

Humidity

h = \frac{RH \cdot p_s}{p}
hWater-vapour fraction
RHRelative humidity as a fraction
p_sSaturation vapour pressure of water at the air temperature in hPa
pAmbient pressure (absolute) in hPa
TemperatureSaturation vapour pressure
0 °C6.1 hPa
10 °C12.3 hPa
20 °C23.4 hPa
25 °C31.7 hPa
30 °C42.5 hPa
40 °C73.8 hPa

Humid air holds less oxygen than dry air. An analyser set to 20.9 % in humid air therefore reads dry cylinder gas too high.

Altitude and weather

p = 1013.25 \cdot (1 - 2.25577 \cdot 10^{-5} \cdot h)^{5.25588}
pAmbient pressure (absolute) in hPa (standard atmosphere)
hAltitude in m
AltitudeAmbient pressure
0 m1,013 hPa
500 m955 hPa
1,000 m899 hPa
1,500 m846 hPa
2,000 m795 hPa
3,000 m701 hPa

The reading changes in proportion to the air pressure. A change of weather of 20 hPa between calibration and measurement shifts the reading by 2 % of its value.

Temperature

Without compensation the cell’s current rises by about 2.5 % for every °C:

G = (1 + \alpha)^{t_m - t_c}
GTemperature factor of an uncompensated cell
\alphaTemperature coefficient as a fraction per °C (0.025 for 2.5 % per °C)
t_mSensor temperature at measurement in °C
t_cSensor temperature at calibration in °C

Nitrox analysers carry a compensation network that cancels most of this; what is left is typically within ±2 % of the reading over 0 to 40 °C. After a change of temperature the cell needs 10 to 30 minutes to settle; gas expanding from a cylinder is cold.

Taking a good reading

  • Calibrate at the place and temperature of the measurement.
  • Calibrate in dry air if possible.
  • Let the gas flow gently, 0.5 to 2 L/min: a strong flow pressurises the cell and reads high.
  • Read a still or slowly flowing sample.