An oxygen analyser doesn’t measure the percentage of oxygen. Its sensor responds to the partial pressure of oxygen, so humidity, altitude, the weather and temperature all move the reading. Calibrate to 20.9 % in humid air, or at home and then use the analyser at a dive site in the mountains, and the same gas shows a different figure.
Enter the gas and the conditions at calibration and at measurement, and the calculator estimates what the analyser should show. Or enter what it shows to find out how much oxygen the gas really holds.
Background
What the cell measures
The cell in a nitrox analyser produces a current in proportion to the partial pressure of oxygen at its membrane, not to the percentage. At calibration the analyser is told which percentage that current means; afterwards it turns the current back into a percentage on that basis.
p_{O_2} = x \cdot (p - p_w)p_{O_2}Partial pressure of oxygen in hPaxOxygen fraction of the dry gaspAmbient pressure (absolute) in hPap_wPartial pressure of water vapour in hPaThe reading
D = S \cdot \frac{x_m \cdot (1 - h_m) \cdot p_m}{x_c \cdot (1 - h_c) \cdot p_c} \cdot GDDisplayed reading in %SValue set at calibration in %x_mOxygen fraction of the dry gas measuredh_mWater-vapour fraction of the gas measuredp_mAmbient pressure at measurement (absolute) in hPax_cOxygen fraction of the dry calibration gas (0.20946 for air)h_cWater-vapour fraction of the calibration airp_cAmbient pressure at calibration (absolute) in hPaGTemperature factor of the cell (1 for a compensated cell)An analyser with a built-in barometer corrects for the air pressure itself; then p_m and p_c drop out. For ‘What is really in it?’ the same formula is solved for x_m.
Humidity
h = \frac{RH \cdot p_s}{p}hWater-vapour fractionRHRelative humidity as a fractionp_sSaturation vapour pressure of water at the air temperature in hPapAmbient pressure (absolute) in hPa| Temperature | Saturation vapour pressure |
|---|---|
| 0 °C | 6.1 hPa |
| 10 °C | 12.3 hPa |
| 20 °C | 23.4 hPa |
| 25 °C | 31.7 hPa |
| 30 °C | 42.5 hPa |
| 40 °C | 73.8 hPa |
Humid air holds less oxygen than dry air. An analyser set to 20.9 % in humid air therefore reads dry cylinder gas too high.
Altitude and weather
p = 1013.25 \cdot (1 - 2.25577 \cdot 10^{-5} \cdot h)^{5.25588}pAmbient pressure (absolute) in hPa (standard atmosphere)hAltitude in m| Altitude | Ambient pressure |
|---|---|
| 0 m | 1,013 hPa |
| 500 m | 955 hPa |
| 1,000 m | 899 hPa |
| 1,500 m | 846 hPa |
| 2,000 m | 795 hPa |
| 3,000 m | 701 hPa |
The reading changes in proportion to the air pressure. A change of weather of 20 hPa between calibration and measurement shifts the reading by 2 % of its value.
Temperature
Without compensation the cell’s current rises by about 2.5 % for every °C:
G = (1 + \alpha)^{t_m - t_c}GTemperature factor of an uncompensated cell\alphaTemperature coefficient as a fraction per °C (0.025 for 2.5 % per °C)t_mSensor temperature at measurement in °Ct_cSensor temperature at calibration in °CNitrox analysers carry a compensation network that cancels most of this; what is left is typically within ±2 % of the reading over 0 to 40 °C. After a change of temperature the cell needs 10 to 30 minutes to settle; gas expanding from a cylinder is cold.
Taking a good reading
- Calibrate at the place and temperature of the measurement.
- Calibrate in dry air if possible.
- Let the gas flow gently, 0.5 to 2 L/min: a strong flow pressurises the cell and reads high.
- Read a still or slowly flowing sample.